An in-house software program for quantitative image quality evaluation of linac cone-beam CT system
Creators
- 1. Advanced Medical and Dental Institute, Universiti Sains Malaysia, Bertam, 13200 Kepala Batas, Pulau Pinang (Malaysia)
Description
The study presents application of an in-house software program to calculate image quality parameters of kV cone-beam CT (CBCT) system in radiotherapy. The parameters include uniformity, noise, contrast-to-noise ratio, CT number linearity, spatial resolution and geometric distortion as recommended in AAPM TG142 and TG179 quality assurance (QA) programme for image quality performance of kV CBCT. The computer algorithm packaged in the software program is developed using MATLAB (MathWorks, Natick, MA). The algorithm was optimised for two CT image quality phantoms, Catphan 600 (The Phantom Laboratory, Salem, NY) and CIRS 062QA (CIRS, Norfolk, VA). The phantoms were scanned with XVI CBCT system (Elekta AB, Stockholm, Sweden) using an optimised CT imaging protocol. The algorithm measures the image quality metrices from the corresponding image quality test patterns in the reconstructed CT images. The algorithm provides image quality parameters from both phantoms quantitatively. The software program allows the results to be exported and recorded in MS Excel for monitoring the performance trend of the image quality parameters. The software reduces measurement uncertainties from qualitative human observation of the test patterns. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/1248/1/012053Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 1248
- Journal Issue
- 1
- Journal Page Range
- [6 p.]
- ISSN
- 1742-6596
Conference
- Title
- 18. Asia-Oceania Congress of Medical Physics (AOCMP); 16. South-East Asia Congress of Medical Physics (SEACOMP)
- Dates
- 11-14 Nov 2018
- Place
- Kuala Lumpur (Malaysia)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53055729
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; COMPUTER CODES; COMPUTERIZED TOMOGRAPHY; GEOMETRY; LINEAR ACCELERATORS; METRICS; MONITORING; PERFORMANCE; PHANTOMS; QUALITY ASSURANCE; RADIOTHERAPY; SPATIAL RESOLUTION
- Descriptors DEC
- ACCELERATORS; DIAGNOSTIC TECHNIQUES; MANAGEMENT; MATHEMATICAL LOGIC; MATHEMATICS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; RESOLUTION; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY