Measurement of off-axis and peripheral skin dose using radiochromic film

  • 1. Illawarra Cancer Care Centre, Department of Medical Physics, Crown Street, Wollongong, NSW 2500 (Australia)
  • 2. City University of Hong Kong, Department of Physics and Materials Science, Kowloon Tong (Hong Kong)

Description

A radiotherapy skin dose profile can be obtained with radiochromic film. The central axis skin dose relative to Dmax for a 10x10cm2 field size was found to be 22%, 17% and 15.5% for 6 MV, 10 MV and 18 MV photon beams. Peripheral dose increased with increasing field size. At 10 MV the skin dose 2 cm outside the geometric field edge was measured as 6%, 10% and 17% for 10x10cm2, 20x20cm2 and 30x30cm2 field sizes respectively. Off-axis skin dose decreased as distance increased from central axis for fields with Perspex block trays. For a 20x20cm2 field, an approximately 5-8% drop in percentage skin dose was observed from central axis to the beam edge. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
43
Journal Issue
9
Journal Page Range
p. 2647-2650
ISSN
1361-6560

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43041973
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
FILM DOSIMETRY; PHOTOGRAPHIC FILM DOSEMETERS; RADIATION DOSES; RADIOTHERAPY; SKIN
Descriptors DEC
BODY; DOSEMETERS; DOSES; DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY

Optional Information

Notes
10 refs; This record replaces 31036352