Measurement of off-axis and peripheral skin dose using radiochromic film
- 1. Illawarra Cancer Care Centre, Department of Medical Physics, Crown Street, Wollongong, NSW 2500 (Australia)
- 2. City University of Hong Kong, Department of Physics and Materials Science, Kowloon Tong (Hong Kong)
Description
A radiotherapy skin dose profile can be obtained with radiochromic film. The central axis skin dose relative to Dmax for a 10x10cm2 field size was found to be 22%, 17% and 15.5% for 6 MV, 10 MV and 18 MV photon beams. Peripheral dose increased with increasing field size. At 10 MV the skin dose 2 cm outside the geometric field edge was measured as 6%, 10% and 17% for 10x10cm2, 20x20cm2 and 30x30cm2 field sizes respectively. Off-axis skin dose decreased as distance increased from central axis for fields with Perspex block trays. For a 20x20cm2 field, an approximately 5-8% drop in percentage skin dose was observed from central axis to the beam edge. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 43
- Journal Issue
- 9
- Journal Page Range
- p. 2647-2650
- ISSN
- 1361-6560
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43041973
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- FILM DOSIMETRY; PHOTOGRAPHIC FILM DOSEMETERS; RADIATION DOSES; RADIOTHERAPY; SKIN
- Descriptors DEC
- BODY; DOSEMETERS; DOSES; DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY
Optional Information
- Notes
- 10 refs; This record replaces 31036352