Published January 2018 | Version v1
Journal article

Using magnetic beads and signal amplifiers to produce short and simple immunoassays: Application to MMP-9 detection in plasma samples

  • 1. Diagnostic Nanotools Group, Cibbim-Nanomedicine, Vall Hebron Research Institute (VHIR), Universitat Autònoma de Barcelona, Barcelona (Spain)
  • 2. Laboratory of Biophysics, Faculty of Medicine of Monastir, University of Monastir (Tunisia)
  • 3. Neurovascular Research Laboratory, Vall d'Hebron Institute of Research (VHIR), Universitat Autònoma de Barcelona, Barcelona (Spain)
  • 4. CIBER de Bioingeniería, Biomateriales y Nanomedicina (CIBER-BBN) (Spain)

Description

Highlights: • Commercial poly-HRP and magnetic beads are used to develop a 35-min magneto-ELISA. • This shortened assay provides results comparable to a classical ELISA that takes >5h. • The shortened assay is finally calibrated by detecting MMP-9 in 12 plasma samples. • The simplified assay is better suited for fast detection than multi-step assays. - Abstract: Magnetic beads (MB) and signal amplifiers, such as horseradish peroxidase polymers (poly-HRP), have been used before for the production of highly sensitive immunoassays. However, most of the examples reported previously entailed long and tedious multi-step procedures, which were not necessarily shorter or simpler than classical paths such as Enzyme-Linked Immunosorbent Assay (ELISA). Here, instead of exploiting the combination of MB and poly-HRP to ameliorate sensitivity, we show that they conform a powerful tool that can be used to shorten the incubation times, which allows optimizing extremely simple, fast and efficient immunoassays with minimal technical requirements. In order to do so, here we used the highly sensitive and specific pair of antibodies of a commercial ELISA kit to optimize a magneto-ELISA for the detection of matrix metallopeptidase 9 (MMP-9). Three signal amplifiers were then tested and the best performing one was implemented in the magneto-assay to shorten the incubation times and improve assay performance. As we show, the shortened magneto-assay could be carried out in about 35 min, which included two 5-min incubations, washing, and incubation with enzyme substrate for 20 min before colorimetric detection. Moreover, the quantification of MMP-9 provided by the shortened assay in 12 plasma samples collected from patients was comparable to that generated by the 5-h ELISA, which was 8.5 times longer.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.aca.2017.11.013

Additional details

Identifiers

DOI
10.1016/j.aca.2017.11.013;
PII
S0003267017312655;

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
999
Journal Page Range
p. 144-154
ISSN
0003-2670
CODEN
ACACAM

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49107127
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
AMPLIFIERS; DETECTION; ENZYME IMMUNOASSAY; HYDROGEN 5; INCUBATION; PLASMA; SENSITIVITY; SIGNALS
Descriptors DEC
BIOASSAY; ELECTRONIC EQUIPMENT; EQUIPMENT; HYDROGEN ISOTOPES; IMMUNOASSAY; ISOTOPES; LIGHT NUCLEI; NUCLEI; ODD-EVEN NUCLEI

Optional Information

Copyright