Published December 2011
| Version v1
Journal article
Improved SNR of magnetic resonance microimaging using a cooled resonance circuit at 0.3T
Creators
- 1. National Inst. of Advanced Industrial Science and Technology, Ikeda, Osaka (Japan)
- 2. National Center for Geriatrics and Gerontology, Obu, Aichi (Japan)
Description
Because it is noninvasive, magnetic resonance microimaging (MRMI) can be used for 3-dimensional measurement of living tissues for cell engineering. Thermal noise in the resonance circuit of the radiofrequency (RF) system of the MRMI cannot become ignored as the signal diminishes in accordance with decreasing sample size, and cooling the RF coil of the receiver circuit can effectively reduce thermal noise. We used a low temperature normal conductor circuit to reduce noise and confirmed improved signal-to-noise ratio for a conventional microimaging system at low B0 field (0.3T) with low cost. (author)
Additional details
Publishing Information
- Journal Title
- Magnetic Resonance in Medical Sciences
- Journal Volume
- 10
- Journal Issue
- 4
- Journal Page Range
- p. 259-262
- ISSN
- 1347-3182
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 43069405
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COOLING; NMR IMAGING; PHANTOMS; RF SYSTEMS; SIGNALS; SIGNAL-TO-NOISE RATIO; SPATIAL RESOLUTION; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; MOCKUP; RESOLUTION; STRUCTURAL MODELS