Published December 1961 | Version v1
Miscellaneous Open

Isothermal calorimeter for reactor radiation dosimetry

  • 1. Institute of Nuclear Sciences Boris Kidric, Odeljenje za radijacionu hemiju, Vinca, Beograd (Serbia)

Description

An isothermal calorimeter with thermistors for measuring absorbed dose rates from 104-5-6.105 rad/h in reactor experimental holes has been designed. A kinetics method for determining the equilibrium temperature difference has been developed, and its application in isothermal calorimetry proved. The expected accuracy in measurements within ± 2-5% has been proved by measurements carried out in the reactor. Some data obtained by measurements in the reactor RA are presented (author)

Files

41124421.pdf

Files (1.1 MB)

Name Size Download all
md5:fe29c3ce5643e8ed7cd9b9fdbd256c7d
1.1 MB Preview Download
Part of:
Calorimeter measurements of absorbed doses at the heavy water enriched uranium reactor

Additional details

Additional titles

Original title (English)
Kalorimetrijska merenja apsorbovanih doza u reaktoru na tesku vodu i obogaceni uran

Publishing Information

Imprint Place
Beograd-Vinca (Serbia)
Imprint Title
Calorimeter measurements of absorbed doses at the heavy water enriched uranium reactor
Imprint Pagination
101 p.
Journal Page Range
19 p.
Report number
INIS-CS--0281

Optional Information

Notes
3 tabs, 6 figs, 8 refs; This record replaces 38039531 Imprint:14 tabs, 30 figs, 15 refs;Kalorimetrijska merenja apsorbovanih doza u reaktoru na tesku vodu i obogaceni uran
Secondary number(s)
IZ--051-0061-1961