The results of measuring characteristics of cryogenic thermometers-dosimeters on a synchrophasotron beam

  • 1. Joint Inst. for Nuclear Research, Dubna (USSR). Lab. of High Energy
  • 2. Moskovskij Inzhenerno-Fizicheskij Inst. (USSR)

Description

Cryogenic thermometers-dosimeters and measurement methods based on radiation heating of thermometers were developed. The experimental installation and construction of detectors cooled in liquid helium is described. The experiment was performed on 2.55 GeV proton and 7.31 GeV deuteron beams at 6 ms and 350 ms pulse duration. The dependences of maximum over heating (up to 7 K) of thermometers from particle fluence are obtained. Coefficients of transition from particle fluence to radiation heating for each thermometers are determined. For 7.31 GeV deuteron beam the coefficients are as follows: for copper-nickel thermometers K=2.6x10-14 Jxcm2/part; for sensors TVO K=2.4x10-14 Jxcm2/part; for sensors ''Allen-Bradley'' (USA) K=4.7x10-15 Jxcm2/part

Availability note (English)

MF available from INIS under the Report Number.

Files

16046597.pdf

Files (168.0 kB)

Name Size Download all
md5:cb7cc3fbc5d85bdced43b0e3fe139d4e
168.0 kB Preview Download

System files (21.6 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Результаты измерения характеристик криогенных термодатчиков-дозиметров на пучке синхрофазотрона

Publishing Information

Imprint Pagination
6 p.
Report number
JINR--9-84-432

Optional Information

Notes
6 refs.; 8 figs.; 1 tab.