Published May 10, 1988
| Version v1
Journal article
Test of an iron/streamer tube calorimeter with electrons and pions of energy between 1 and 100 GeV
- 1. Siegen Univ. (Gesamthochschule) (Germany, F.R.). Fachbereich 7 - Physik
Description
We have constructed a simple and low cost modular electron/hadron streamer tube calorimeter with 2 cm (resp. 4 cm) iron sampling. To suppress saturation effects we use small cell streamer tubes (6 mm x 6 mm) filled with pure isobutane. The calorimeter was tested with electrons and pions in the energy range from 1 to 100 GeV. The linearity of the electron response extends to about 25 GeV. The response to pions per GeV increases at the highest energies by about 20%. The energy resolution for electrons (pions) is determined to be σ/E = 3.8% + (28.8%/√E)(σ/E = 3.9% + (72.1%/√E)), where the hadron resolution can be improved considerably by a weighting procedure for individual channels. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 268
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 105-111
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 19069088
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALORIMETERS; ELECTRON DETECTION; ENERGY DEPENDENCE; ENERGY RESOLUTION; GEV RANGE 01-10; GEV RANGE 10-100; IRON; PION DETECTION; RELATIVISTIC RANGE; RESPONSE FUNCTIONS; STREAMER SPARK CHAMBERS; TUBES; 2-METHYLPROPANE
- Descriptors DEC
- ALKANES; CHARGED PARTICLE DETECTION; DETECTION; ELEMENTS; ENERGY RANGE; FUNCTIONS; GAS TRACK DETECTORS; GEV RANGE; HYDROCARBONS; MEASURING INSTRUMENTS; METALS; ORGANIC COMPOUNDS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SPARK CHAMBERS; TRANSITION ELEMENTS