Studies of a proton phase beam monitor for range verification in proton therapy
Creators
- 1. Technische Universitaet Dresden, OncoRay, PF 41, 01307 Dresden, (Germany)
- 2. Technische Universitaet Dresden, Institute for Nuclear and Particle Physics, Zellescher Weg 19, 01069 Dresden, (Germany)
- 3. Helmholtz-Zentrum Dresden-Rossendorf, PF 510 119, 01314 Dresden, (Germany)
Description
A primary subject of the present research in particle therapy is to ensure the precise irradiation of the target volume. The prompt gamma timing (PGT) method provides one possibility for in vivo range verification during the irradiation of patients. Prompt gamma rays with high energies are emitted promptly due to nuclear reactions of protons with tissue. The arrival time of these gammas to the detector reflects the stopping process of the primary protons in tissue and are directly correlated to the range. Due to the time resolution of the detector and the proton bunch time spread, as well as drifts of the bunch phase with respect to the accelerator frequency, timing spectra are smeared out and compromise the accuracy of range information intended for future clinical applications. Nevertheless, counteracting this limitation and recovering range information from the PGT measured spectra, corrections using a phase beam monitor can be performed. A first prototype of phase beam monitor was tested at GSI Darmstadt, where measurements of the energy profile of the ion bunches were performed. At the ELBE accelerator Helmholtz-Zentrum Dresden-Rossendorf (HZDR), set up to provide bremsstrahlung photons in very short pulses, a constant fraction algorithm for the incoming digital signals was evaluated, which is used for optimizing the time resolution. Studies of scattering experiments with different thin targets and detector positions are accomplished at Oncoray Dresden, where a clinical proton beam is available. These experiments allow a basic characterization of the proton bunch structure and the detection yield. (authors)
Additional details
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- ANIMMA--2015-IO-172
Conference
- Title
- 4. International Conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications
- Acronym
- ANIMMA 2015
- Dates
- 20-24 Apr 2015
- Place
- Lisboa (Portugal)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- France
- INIS RN
- 47102167
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ACCELERATORS; ALGORITHMS; ANIMAL TISSUES; BREMSSTRAHLUNG; GAMMA RADIATION; IN VIVO; IRRADIATION; NUCLEAR REACTIONS; OPTIMIZATION; PATIENTS; PHOTONS; PROTON BEAMS; PROTONS; RADIOTHERAPY; TIME RESOLUTION; VERIFICATION
- Descriptors DEC
- BARYONS; BEAMS; BODY; BOSONS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; HADRONS; IONIZING RADIATIONS; MASSLESS PARTICLES; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; NUCLEON BEAMS; NUCLEONS; PARTICLE BEAMS; RADIATIONS; RADIOLOGY; RESOLUTION; THERAPY; TIMING PROPERTIES
Optional Information
- Notes
- 9 Refs.