Preliminary study plate capacitor as a Plethysmometer sensor
Description
In this research, we successfully designed a capacitor plate as a Plethysmometer sensor with a measured volume capacity of 1.5 ml, this capacity is obtained from a capacitor plate with a thickness of 0.5 mm width of 15 mm and a height of 200 mm. The sensitivity obtained is 0.36 / 0.1ml with a standard deviation of 0.012. The above results show that the linear capacitor plate sensor for the tested area has sufficient sensitivity that can measure the change of 0.1 ml to be used as a transducer in the Plethysmometer. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1757-899X/403/1/012036Additional details
Identifiers
Publishing Information
- Journal Title
- IOP Conference Series. Materials Science and Engineering (Online)
- Journal Volume
- 403
- Journal Issue
- 1
- Journal Page Range
- [5 p.]
- ISSN
- 1757-899X
Conference
- Title
- 1. International Conference on Engineering and Applied Technology (ICEAT)
- Dates
- 29-30 Nov 2017
- Place
- Mataram (Indonesia)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52095006
- Subject category
- S47: OTHER INSTRUMENTATION; S42: ENGINEERING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CAPACITORS; CAPACITY; DESIGN; PLATES; SENSITIVITY; SENSORS; TRANSDUCERS
- Descriptors DEC
- ELECTRICAL EQUIPMENT; EQUIPMENT