Published April 2006 | Version v1
Journal article

Radiation doses to patients receiving computed tomography examinations in British Columbia

  • 1. Univ. of British Columbia, Vancouver Hospital, Dept. of Radiology, Vancouver, British Columbia (Canada)

Description

To estimate the diagnostic reference levels and effective radiation dose to patients from routine computed tomography (CT) examinations in the province of British Columbia, Canada. The patient weight, height and computed tomography dose index or dose linear product (DLP) were recorded on study sheets for 1070 patients who were referred for clinically indicated routine CT examinations at 18 radiology departments in British Columbia. Sixteen of the scanners were multidetector row scanners. The average patient dose varied from hospital to hospital. The largest range was found for CT of the abdomen, for which the dose varied from 3.6 to 26.5 (average 10.1) mSv. For head CT, the range was 1.7 to 4.9 (average 2.8) mSv; for chest CT, it was 3.8 to 26 (average 9.3) mSv; for pelvis CT, it was 3.5 to 15.5 (average 9.0) mSv; and for abdomen/pelvis CT, it was 7.3 to 31.5 (average 16.3) mSv. Reference dose values were calculated for each exam. These DLP values are as follows: head, 1300 mGy cm; chest, 600 mGy cm; abdomen, 920 mGy cm; pelvis, 650 mGy cm; and abdomen/pelvis, 1100 mGy cm. Among hospitals, there was considerable variation in the DLP and patient radiation dose for a specific exam. Reference doses and patient doses were higher than those found in similar recent surveys carried out in the United Kingdom and the European Union. Patient doses were similar to those found in a recent survey in Germany. (author)

Additional details

Publishing Information

Journal Title
Canadian Association of Radiologists Journal
Journal Volume
57
Journal Issue
2
Journal Page Range
p. 79-85
ISSN
0846-5371

INIS

Country of Publication
Canada
Country of Input or Organization
Canada
INIS RN
38036447
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ABDOMEN; BRITISH COLUMBIA; COMPUTERIZED TOMOGRAPHY; IMAGE SCANNERS; PARAMETRIC ANALYSIS; PATIENTS; PELVIS; RADIATION DOSES
Descriptors DEC
BODY; CANADA; DEVELOPED COUNTRIES; DIAGNOSTIC TECHNIQUES; DOSES; NORTH AMERICA; TOMOGRAPHY

Optional Information

Notes
22 refs., 6 tabs., 5 figs.