A simple ion energy analyser equipped with a quadrupole mass spectrometer
- 1. Ulvac Corp., Chigasaki, Kanagawa (Japan)
Description
A cylindrical mirror analyser to pass ions in the energy range from 0 to 10 eV and at the angle of incidence ranging from 70 to 110 is designed for a co-axial ion energy analyser (IEA) suitable for combining with a quadrupole mass spectrometer (QMS) for secondary ion mass spectrometry (SIMS). A mass resolution of the combination for Cr+ in SIMS measurement shows approximately that of the QMS in residual gas analysis. An ion detector using the combination to measure the energy distribution of sputtered secondary ions is calibrated with a Na+ ion gun. The energy resolution and the accuracy are within 5 eV and 0.5 eV, in the ion energy range of 5 to 60 eV respectively. Energy distributions of various metal ions in including Na+, K+, Cr+, Fe+, Al+, Cu+, Ni+, Ti+, and Si+ are measured under Xe ion bombardment of 0.5 to 2 keV. These are compared with energy distributions measured by other investigators. (auth.)
Additional details
Identifiers
- DOI
- 10.1143/jjap.15.1359;
Publishing Information
- Journal Title
- Japanese Journal of Applied Physics
- Journal Volume
- 15
- Journal Issue
- 7
- Series
- Jpn. J. Appl. Phys.
- Journal Page Range
- 1359-1366
- ISSN
- 0021-4922
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 8307458
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; BEAM DYNAMICS; CALIBRATION; ENERGY RESOLUTION; EV RANGE 01-10; ION BEAMS; ION SPECTROSCOPY; IONS; MAGNETIC FIELDS; MASS RESOLUTION; MASS SPECTROMETERS; SPATIAL DISTRIBUTION
- Descriptors DEC
- BEAMS; CHARGED PARTICLES; DISTRIBUTION; ENERGY RANGE; EV RANGE; MEASURING INSTRUMENTS; RESOLUTION; SPECTROMETERS; SPECTROSCOPY
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent