Published February 2010 | Version v1
Journal article

3D dose delivery verification using repeated cone-beam imaging and EPID dosimetry for stereotactic body radiotherapy of non-small cell lung cancer

  • 1. Department of Radiation Oncology (MAASTRO), Maastricht University Medical Centre (Netherlands)

Description

Purpose: To implement a 3D dose verification procedure, based on in-room cone-beam CT imaging and portal dosimetry, for lung cancer patients treated with stereotactic body radiotherapy (SBRT). Materials and methods: MV cone-beam CT scans were made for patient positioning and calibrated for dose calculation purposes. Prior to treatment, the treatment fields were captured using a calibrated electronic portal imaging device (EPID). A Monte Carlo dose reconstruction model was used to estimate the 3D dose delivered to the patient inside the cone-beam CT images. The planned and delivered dose distributions were compared for 4 patients and 10 treatment fractions using dose-volume histograms and gamma analysis. Results: The gamma analysis showed a good agreement between the planned and delivered dose distributions for patients without changes in anatomy. The delivered mean dose per fraction inside the target volume deviated on average 1.1 ± 1.4% from the planned dose. For the critical organs, only minor differences were observed between the reconstructed and planned dose. Conclusions: A method was presented that allows verification of the dose delivered in 3D for lung cancer patients treated with SBRT. The procedure is independent of the treatment planning system and uses in-room MV cone-beam CT imaging and portal dosimetry.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radonc.2009.12.024

Additional details

Identifiers

DOI
10.1016/j.radonc.2009.12.024;
PII
S0167-8140(09)00682-3;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
94
Journal Issue
2
Journal Page Range
p. 188-194
ISSN
0167-8140
CODEN
RAONDT

Conference

Title
10. biennial ESTRO conference on physics and radiation technology for clinical radiotherapy
Dates
30 Aug - 3 Sep 2009
Place
Maastricht (Netherlands)

INIS

Country of Publication
Ireland
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41074389
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CAT SCANNING; CRITICAL ORGANS; DOSIMETRY; LUNGS; MONTE CARLO METHOD; NEOPLASMS; PATIENTS; PLANNING; POSITIONING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; VERIFICATION
Descriptors DEC
BODY; CALCULATION METHODS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESPIRATORY SYSTEM; THERAPY; TOMOGRAPHY