Published May 15, 1985 | Version v1
Journal article

Trace element analysis by using the direct CPAA method

  • 1. Magyar Tudomanyos Akademia, Debrecen. Atommag Kutato Intezete

Description

Multielemental trace analysis was performed at Turku Cyclotron Laboratory on purified industrial aluminium samples and Ultra-High-Purity aluminium standards, produced by VAW (FRG), by using the direct CPAA (Charged Particle Activation Analysis) method. Both types of Al were expected to contain less than 10 ppm trace element contamination. 8 to 10 MeV proton beams were used for irradiation. The absolute trace quantities were determined for some ten elements by using the cross section curves and calculating the ranges of the bombarding particles in the samples, and measuring the absolute beam current and the intensities of the characteristic gamma-lines of the activated isotopes. The investigated elements are: Ca, Ti, V, Cr, Fe, Ni, Cu, Zn, Ga, Br and Zr. The obtained results were in good agreement with the guaranteed (VAW) data. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. B
Journal Volume
10/11
Journal Issue
pt.2
Series
CODEN: NIMBE.;Nucl. Instrum. Methods Phys. Res., Sect. B.
Journal Page Range
1051-1053
ISSN
0168-583X

Conference

Title
8. international conference on the application of accelerators in research and industry.
Dates
12-14 Nov 1984.
Place
Denton, TX (USA).