Published May 2010 | Version v1
Journal article

High frequency, high time resolution time-to-digital converter employing passive resonating circuits

  • 1. Department of Electronics, Politecnico di Milano, via C. Golgi 40, Milano 20133 (Italy)

Description

A method for measuring time intervals accurate to the picosecond range is based on phase measurements of oscillating waveforms synchronous with their beginning and/or end. The oscillation is generated by triggering an LC resonant circuit, whose capacitance is precharged. By using high Q resonators and a final active quenching of the oscillation, it is possible to conjugate high time resolution and a small measurement time, which allows a high measurement rate. Methods for fast analysis of the data are considered and discussed with reference to computing resource requirements, speed, and accuracy. Experimental tests show the feasibility of the method and a time accuracy better than 4 ps rms. Methods aimed at further reducing hardware resources are finally discussed.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
81
Journal Issue
5
Journal Page Range
p. 054705-054705.7
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2010 American Institute of Physics