Description

The influence of the adsorbed mass of molybdenum on the width of eluation profiles of generators and the patterns of molybdenum distribution in the amount of chromatographic columns by scanning them on the germanium-gallium detector using collimating device are studied. The boundary conditions under which the maximum value of 99mTc yield from generators Ye = 1 are defined. After scanning the columns, it was found out that the degree of filling the columns with molybdenum Q and the value of its maximum adsorption depend naturally on the total weight of the adsorbed mass. In order to achieve the condition Ye = 1 the value of Q should be at least 85%.