Aortic annulus eccentricity before and after transcatheter aortic valve implantation: Comparison of balloon-expandable and self-expanding prostheses

  • 1. Department of Cardiology, University of Erlangen, Erlangen (Germany)
  • 2. Department of Cardiology, University of Gießen, Giessen (Germany)
  • 3. Department of Radiology, University of Gießen, Giessen (Germany)
  • 4. Department of Radiology, University of Erlangen, Erlangen (Germany)
  • 5. Department of Thoracic and Cardiovascular Surgery, Heart and Diabetes Center NRW, Ruhr-University Bochum, Bad Oeynhausen (Germany)
  • 6. Department of Cardiothoracic Surgery, University of Erlangen, Erlangen (Germany)

Description

Highlights: • Post-implant geometry of catheter-based aortic valve prostheses is influenced by aortic valve calcification. • Balloon-expandable prostheses are more circular as compared to self-expanding prostheses. • The impact of post-implant geometry on valve function needs to be investigated. - Abstract: Introduction: The geometry of the aortic annulus and implanted transcatheter aortic valve prosthesis might influence valve function. We investigated the influence of valve type and aortic valve calcification on post-implant geometry of catheter-based aortic valve prostheses. Methods: Eighty consecutive patients with severe aortic valve stenosis (mean age 82 ± 6 years) underwent computed tomography before and after TAVI. Aortic annulus diameters were determined. Influence of prosthesis type and degree of aortic valve calcification on post-implant eccentricity were analysed. Results: Aortic annulus eccentricity was reduced in patients after TAVI (0.21 ± 0.06 vs. 0.08 ± 0.06, p < 0.0001). Post-TAVI eccentricity was significantly lower in 65 patients following implantation of a balloon-expandable prosthesis as compared to 15 patients who received a self-expanding prosthesis (0.06 ± 0.05 vs. 0.15 ± 0.07, p < 0.0001), even though the extent of aortic valve calcification was not different. After TAVI, patients with a higher calcium amount retained a significantly higher eccentricity compared to patients with lower amounts of calcium. Conclusions: Patients undergoing TAVI with a balloon-expandable prosthesis show a more circular shape of the implanted prosthesis as compared to patients with a self-expanding prosthesis. Eccentricity of the deployed prosthesis is affected by the extent of aortic valve calcification

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2015.04.003

Additional details

Identifiers

DOI
10.1016/j.ejrad.2015.04.003;
PII
S0720-048X(15)00177-1;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
84
Journal Issue
7
Journal Page Range
p. 1242-1248