Published June 2009 | Version v1
Journal article

Preparation of 65Zn and 134Cs sealed sources based on commercial organic and synthesized inorganic sorbents for calibration purposes of γ-detectors

  • 1. Radioactive Isotopes and Generators Department, Hot Lab Center, Atomic Energy Authority, PO Box 13759, Cairo (Egypt)

Description

The sealed source of 65Zn was prepared by using permutit as a commercial organic resin in an anion form, while in case of the 134Cs, the sealed source was prepared by using the synthesized 12-molybdocerate(IV) gel as an inorganic exchanger in cation form. The batch technique was carried out for loading 65Zn from 0.2 M HCl-60% acetone medium and for loading 134Cs from 0.005 M HNO3 onto the appropriate exchanger. The sealed source material of 65Zn-permutit or 134Cs-12-molybdocerate(IV), about 100 mg, was placed inside the cavity of a holder capsule (artelone) and its activity was measured to be 56.61±2.59 kBq (1.53±0.07 μCi) for 65Zn on 21 April 2008 and 1.48±0.089 kBq (0.04±0.0024 μCi) for 134Cs on 5 May 2008. The self absorption was calculated by using artelone and compared with the same thickness of stainless steel capsule as the holder material. The quality control was performed on the prepared sealed sources.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2008.12.017

Additional details

Identifiers

DOI
10.1016/j.apradiso.2008.12.017;
PII
S0969-8043(09)00002-5;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
67
Journal Issue
6
Journal Page Range
p. 984-989
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2009 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.