Published May 26, 2008 | Version v1
Journal article

Evaluation of B/A nonlinear parameter using an acoustic self-calibrated pulse-echo method

  • 1. Universite Francois Rabelais de Tours, INSERM U930/CNRS FRE2448/IC-CUP, ENI Val de Loire, Rue de la Chocolaterie, BP 3410, 41034 Blois Cedex (France)

Description

The objective of this work is to develop an easy-to-build and robust setup for measuring the nonlinearity parameter B/A in fluids using ultrasound. The method is based on the pulse-echo technique, using a single element broadband acoustic transducer, and requires electrical signal measurements. Results obtained in water and denatured alcohol validate the proposed procedure. The choice of a suitable primary wave frequency is discussed with regard to the transducer sensitivity. Further, the influence of the perturbations introduced by the experimental device nonlinearities, and the role of the reflector on the measured second harmonic field amplitude are investigated

Additional details

Identifiers

Publishing Information

Journal Title
Applied Physics Letters
Journal Volume
92
Journal Issue
21
Journal Page Range
p. 214106-214106.3
ISSN
0003-6951
CODEN
APPLAB

Optional Information

Notes
(c) 2008 American Institute of Physics