Published 2000 | Version v1
Computer medium

Study of the characteristics of portable monitors of gamma radiation

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN), Sao Paulo, SP (Brazil)

Description

A series of type tests recommended by international standards was applied to the gamma radiation monitoring survey meters (Geiger-Mueller type, ionization chambers and a scintillator) commonly used in Brazil. The objective was to establish a routine test methodology to be applied to new and to repaired instruments. The tests included the verification of energy and angular dependences, variation of response with the battery voltage and intrinsic errors. (author)

Part of:
Nuclear future: thinking for building. Proceedings of the 5. Brazilian national meeting on nuclear applications; 8. General congress on nuclear energy; 12. Brazilian national meeting on reactor physics and thermal hydraulics

Additional details

Additional titles

Original title (Portuguese)
Estudo das caracteristicas de monitores portateis de radiacao gama

Publishing Information

Imprint Title
Nuclear future: thinking for building. Proceedings of the 5. Brazilian national meeting on nuclear applications; 8. General congress on nuclear energy; 12. Brazilian national meeting on reactor physics and thermal hydraulics
Imprint Pagination
[1450 p.]
Journal Page Range
[4 p.]

Conference

Title
5. Brazilian national meeting on nuclear applications; 8. General congress on nuclear energy; 12. Brazilian national meeting on reactor physics and thermal hydraulics
Original Conference Title
5. Encontro nacional de aplicacoes nucleares
Dates
5-20 Oct 2000
Place
Rio de Janeiro, RJ (Brazil)

Optional Information

Notes
4 refs., 5 tabs. Imprint:Published only in CD-Rom;Futuro nuclear: refletindo para construir. Anais do 5. Encontro nacional de aplicacoes nucleares; 8. Congresso geral de energia nuclear; 12. Encontro nacional de fisica de reatores e termo-hidraulica