Published November 1979 | Version v1
Journal article

An energy correction of TLD (MSO-S type)

  • 1. Nagoya Univ. (Japan). Faculty of Medicine

Description

The detector or thermoluminescence dosimeter shows various sensitivity according to the radiation energy. A detector with MSO type (Mg2SiO4:Tb) of the thermoluminescent material reported in it's different characteristics by YAMAMOTO, HASIZUME and others. This paper reported the method of the measurement for the various kinds of the radiation energy with thermoluminescence dosimeter. At the same time the other thermoluminescent material was used for comparison and obtained the correction factor respectively. This method is applied in the medical field, such as in cases of Radium intracavitary therapy for uterine cancer. (author)

Additional details

Additional titles

Subtitle (English)
A method of energy correction factor for "2"2"6Ra gamma-ray

Publishing Information

Journal Title
Ikakikaigaku Zasshi
Journal Volume
49
Journal Issue
11
Series
Ikakikaigaku Zasshi.
Journal Page Range
515-518
ISSN
0019-1736