Published 2005 | Version v1
Journal article

The DIMOND project and its impact on radiation protection

Creators

  • 1. Quality Assurance Reference Centre, Kingfisher Way, Wallsend, Tyne and Wear NE28 9ND (United Kingdom)

Description

The DIMOND III project comprised work packages and sub-projects to research digital imaging, interventional radiology and interventional cardiology, i.e. areas where there has been rapid technological change, not matched by radiation protection research. Whilst new practices have great potential benefit, there are risks to patients and staff if the technology is implemented inappropriately. DIMOND aimed to develop generic technology assessment and optimisation tools. Clinical quality criteria and technical parameters were included so that digital imaging procedures can be introduced appropriately. Frequency and dimensions of acceptance and constancy testing for digital imaging systems were studied, along with differences between objective measures of image quality and subjective indices. Patient dosimetry was performed to propose and to establish reference values. Clinical evaluation projects included cardiology, interventional radiology and digital mammography. As a result of the DIMOND project, quality criteria for several procedures were developed. Example results are presented in this paper. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci700

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
117
Journal Issue
1-3
Journal Page Range
p. 3-6
ISSN
0144-8420

Conference

Title
3. Conference on Optimisation of dose and performance in interventional and digital imaging
Dates
25-27 Mar 2004
Place
Leuven (Belgium)

Optional Information

Notes
14 refs