Published September 1992 | Version v1
Journal article

Model Bh 1227 4-channel low background α and β measuring assembly

  • 1. Beijing Nurlear Instrument Factory (China)
  • 2. Academia Sinica, Beijing, BJ (China). Inst. of High Energy Physics

Description

The authors present and α and β radiation system of counting emission from up four samples simultaneously. The primary detectors are made of the new low background α and β scintillator of model ST 1221 and the low noise PM tube model GDB-52LD. A single 300 mm x 50 mm plastic scintillator is used as guard detector. 90Sr-90Y counting efficiency is 60% with typical background of 0.03/cm2 · min, counting efficiency for 239Pu is 70% with background of 0.03/cm2· h. This instrument is compatible with microcomputer of any type. Measuring and data processing are controlled by computer

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
12
Journal Issue
suppl.
Series
Proceedings of 6th national coference on nuclear electronics and nuclear detection technology (A).
Journal Page Range
p. 342-346, 351.
ISSN
0258-0934
CODEN
HDYUEC

Conference

Title
6. national conference on nuclear electronics and nuclear detection technology.
Dates
21-26 Sep 1992.
Place
Weihai (China).