Published July 1998 | Version v1
Journal article

Caesium-137 manual after-loading system - calculation of dose rates and their comparison with BRIT supplied values

Creators

  • 1. Manipal Goa Cancer and General Hospital, Don Paula (India)

Description

The new BRIT supplied 137Cs kit contains 3 sources (120 mCi, 80 mCi, 80 mCi) in the long intrauterine tandem, two sources (120 mCi and 80 mCi) in the medium intrauterine tandem and one source (80 mCi) in each vaginal ovoid. Use of sievert-integral can be made to calculate the dose rate at point A and point B. It is concluded that the method gives reliable results for calculating the dose rates at point A and for point B one can reduce the calculated dose rates by around 10 percent

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
23
Journal Issue
3
Journal Page Range
p. 121-123
ISSN
0971-6203

Conference

Title
International conference on medical physics; 19. annual conference on medical physics
Dates
6-9 Nov 1998
Place
New Delhi (India)

INIS

Optional Information

Notes