Published 1998 | Version v1
Journal article

MTF and DQE measurement in imaging detectors by their single-event response

  • 1. Dipartimento di Fisica dell'Universita di Genova and Istituto Nazionale di Fisica Nucleare, Sezione di Genova, via Dodecaneso 33, 16146 Genova (Italy)
  • 2. Dipartimento di Fisica dell'Universita di Padova and Istituto Nazionale di Fisica Nucleare, Sezione di Padova, via Marzolo 8, 35131 Padova (Italy)

Description

The single-event response of imaging detectors, averaged and normalized, can furnish the PSF and then the MTF of the detector. Moreover, the analysis of the single-event responses makes it possible to measure the fluctuation of the output signal intensity and to express the output noises in terms of quantum responses. Therefore, if the variances of the involved noises, the quantum efficiency QE and the average input signal are known, the DQE of the detector can be determined. Experimental results are shown using a CCD-based X-ray scintillation area detector in the 8-45 keV range. (author)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
419
Journal Issue
Nos.2,3
Journal Page Range
p. 731-735
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
8. International Wire Chamber Conference
Dates
23-27 Feb 1998
Place