The measurement with the Medipix2 detector ad the RW3 moderator in the neutron field of the Clinac 2100 linear accelerator
Creators
- 1. Institute of Experimental and Applied Physics, Czech Technical University in Prague, CZ 12800 Prague 2 and Department of Radiological Physics, University hospital na Bulovce, CZ 180 81 Prague 8 (Czech Republic)
- 2. Department of Physics, Faculty of Mechanical Engineering, Czech Technical University in Prague, CZ 16000 Prague 6 (Czech Republic)
- 3. Department of Radiological Physics, University hospital na Bulovce, CZ 180 81 Prague 8 (Czech Republic)
Description
Detection of neutrons in radiotherapy is important to determine the radiation exposure of the patient. Today, the most commonly are used TLD detectors for routine determination of patient dose. These detectors are not accurate. Other applications of Bonner spheres, which is positioned as proportional counter. The detection of neutrons is also possible with Medipix2 detector. Measured results show that using Medipix2 detectors can detect fast neutrons. Fast neutrons are slowed down in the moderator RW3 to thermal neutrons. These neutrons are converted by the LiF6 to charged particles, which are registered by Medipix2 detector. Results of our measurements show the greatest efficiency is achieved with RW3 moderator thickness 40 mm. In the future we plan to place the detector Medipix2 Bonner sphere. This requires a reduction of ambient electronic detector. On this issue has been working intensively. Then we compared the individual responses to detected neutrons which are moderated by Bonner spheres with no available relative dependencies. This principle of neutron detection with detector Medipix2 will be used for monitoring the neutron field in the radiotherapy clinic at The University Hospital Bulovka as part of radiation protection of staff and patients. (authors)
Files
47127129.pdf
Files
(756.5 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:13f03c33cd9ad3210e726a2c5e6da57c
|
756.5 kB | Preview Download |
Additional details
Identifiers
Publishing Information
- Publisher
- University of Zilina
- Imprint Place
- Zilina (Slovakia)
- ISBN
- 978-80-554-0386-1
- Imprint Title
- Proceedings 17. International Conference on Applied Physics of Condensed Matter
- Imprint Pagination
- 328 p.
- Journal Page Range
- p. 231-235
- Report number
- INIS-SK--2016-043
Conference
- Title
- 17. International Conference on Applied Physics of Condensed Matter
- Acronym
- APCOM 2011
- Dates
- 22-24 Jun 2011
- Place
- High Tatras (Slovakia)
INIS
- Country of Publication
- Slovakia
- Country of Input or Organization
- Slovakia
- INIS RN
- 47127129
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference, Numerical Data
- Descriptors DEI
- EXPERIMENTAL DATA; FAST NEUTRONS; LINEAR ACCELERATORS; MODERATORS; NEUTRON DETECTION; NEUTRON DETECTORS; NEUTRON DOSIMETRY; NEUTRON MONITORS; NUCLEAR PHYSICS; RADIOTHERAPY
- Descriptors DEC
- ACCELERATORS; BARYONS; DATA; DETECTION; DOSIMETRY; ELEMENTARY PARTICLES; FERMIONS; HADRONS; INFORMATION; MEASURING INSTRUMENTS; MEDICINE; MONITORS; NEUTRONS; NUCLEAR MEDICINE; NUCLEONS; NUMERICAL DATA; PHYSICS; RADIATION DETECTION; RADIATION DETECTORS; RADIATION MONITORS; RADIOLOGY; THERAPY
Optional Information
- Contract/Grant/Project number
- Project VZ MSM 210000018
- Notes