Published January 1, 2002 | Version v1
Journal article

High-energy neutron reference fields for the calibration of detectors used in neutron spectrometry

Description

Quasi-monoenergetic reference neutron beams in the energy range between 20 and 100 MeV have been produced and characterized with a proton recoil telescope, a scintillation spectrometer, a 238U fission chamber and a Bonner sphere spectrometer. The beams are well suited for the calibration of detectors used in neutron spectrometry. A new method is described which reduces the correction for the contribution from low-energy neutrons present in the beams

Additional details

Identifiers

PII
S0168900201014723;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
476
Journal Issue
1-2
Journal Page Range
p. 369-373
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.