Published June 2015
| Version v1
Journal article
Note: On the wavelength dependence of the intensity calibration factor of extreme ultraviolet spectrometer determined with profile measurement of bremsstrahlung continuum
- 1. Graduate School of Medicine and Pharmaceutical Science, University of Toyama, Toyama 930-0193 (Japan)
- 2. National Institute for Fusion Science, Toki, Gifu 509-5292 (Japan)
- 3. High Energy Accelerator Research Organization, Tsukuba, Ibaraki 305-0801 (Japan)
Description
The absolute calibration factor of extreme ultraviolet spectroscopic instrument which has recently been determined from absolute radiation profile measurement of bremsstrahlung continuum has been investigated by comparing the calculated diffraction efficiency of grating. An overall tendency of the wavelength dependence of the calibration factor from 40 Å to 500 Å can be reproduced by that of the grating efficiency, especially the agreement between the measured calibration factor and the calculated grating efficiency has been found to be fairly good for the wavelength range 200 Å-500 Å
Additional details
Identifiers
- DOI
- 10.1063/1.4922532;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 86
- Journal Issue
- 6
- Journal Page Range
- p. 066105-066105.3
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47053231
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BREMSSTRAHLUNG; CALIBRATION; DIFFRACTION; DIFFRACTION GRATINGS; EFFICIENCY; EXTREME ULTRAVIOLET RADIATION; FREQUENCY DEPENDENCE; SPECTROMETERS; WAVELENGTHS
- Descriptors DEC
- COHERENT SCATTERING; ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATIONS; SCATTERING; ULTRAVIOLET RADIATION
Optional Information
- Notes
- (c) 2015 AIP Publishing LLC