Published September 2013 | Version v1
Journal article

Preparation of cholesterol oxidase nanoparticles and their application in amperometric determination of cholesterol

  • 1. M. D. University, Department of Biochemistry (India)

Description

The nanoparticle (NP) aggregates of commercial cholesterol oxidase (ChOx) were prepared by desolvation method. The formation and characterization of ChOxNP aggregates were studied by transmission electron microscopy and scanning electron microscopy. NP aggregates were more stable, active and had a higher shelf life than that of free enzyme. An amperometric cholesterol biosensor was constructed by immobilizing ChOxNPs onto Au electrode. The biosensor showed optimum response within 8 s at pH 6.0 and 35 °C, when polarized at +0.27 V versus Ag/AgCl. The biosensor possesses high sensitivity and measures cholesterol concentrations as low as 1.56 mg/dl. The working linear range was 12.5–700 mg/dl for cholesterol. The biosensor was evaluated and employed for measurement of total cholesterol in human serum. The enzyme electrode lost 50 % of its initial activity during its regular use for 180 times over a period of 90 days when stored in 0.1 M sodium phosphate buffer, pH 7.0 at 4 °C

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Nanoparticle Research
Journal Volume
15
Journal Issue
9
Journal Page Range
p. 1-9
ISSN
1388-0764

Optional Information

Copyright
Copyright (c) 2013 Springer Science+Business Media Dordrecht