Immune and non-immune T4 radioassays utilizing albumin magnetic microparticles

Description

Albumin magnetic microparticles reversibly adsorb thyroxine. They quickly establish equilibrium allowing time and temperature independent measurements in T4 radioassays. These particles were used to compare the efficiency of NaOH, HCl, pepsin, sodium trichloroacetate, and 8-anilino-1-naphthalene sulfonic acid to release 125I-T4 from serum, T4-free serum and human serum albumin. It was found that the efficiency of the reagents to extract 125I-T4 depended on the concentration and type of proteins to which the labelled hormone was bound. Pepsin was the most effective reagent and it was utilized for a T4 radioimmunoassay, in which albumin magnetic microparticles were used to separate free from bound hormone, the authors also utilized the particles in a T4 non-immune radioassay. Both assays accurately measured total serum T4, however the radioimmunoassay was simpler, less dependent on protein content of serum, required a smaller serum sample and provided slightly higher T4 values. A magnetic rack is described which allows simultaneous handling of fifty individual tubes with an intra-assay C.V. of 2.1% for the radioimmunoassay and 2.3% for the non-immune assay and an inter-assay C.V. of 3.1%, respectively. (Auth.)

Additional details

Publishing Information

Journal Title
Clin. Chim. Acta
Journal Volume
84
Journal Issue
1-2
Series
Clin. Chim. Acta.
Journal Page Range
69-84
ISSN
0009-8981