Dose evaluation for periapical radiographs in the city of Limeira, Sao Paulo, Brazil

  • 1. Vigilancia Sanitaria de Limeira, SP (Brazil)
  • 2. Universidade de Sao Paulo (USP), Ribeirao Preto, SP (Brazil). Faculdade de Filosofia, Ciencias e Letras. Dept. de Fisica e Matematica
  • 3. Universidade de Sao Paulo (USP), Ribeirao Preto, SP (Brazil). Faculdade de Medicina. Hospital das Clinicas

Description

In this work a study of the entrance surface air kerma for periapical radiographs in the city of Limeira, Sao Paulo state, was carried out. Measurements were performed in ten X-ray equipment for periapical radiographs. The values obtained were compared with the diagnostic reference level recommended in national regulation and has suggested a local diagnostic reference level of 2.70 mGy. (author)

Part of:
Proceedings of the 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging

Additional details

Additional titles

Original title (Portuguese)
Avaliacao de dose para radiografias periapicais na cidade de Limeira

Identifiers

Publishing Information

Publisher
ABFM
Imprint Place
Sao Paulo (Brazil)
ISBN
978-85-85205-94-2
Imprint Title
Proceedings of the 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
Imprint Pagination
[vp.]
Journal Page Range
[3 p.]

Conference

Title
14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
Original Conference Title
14. Congresso brasileiro de fisica medica
Dates
8-12 Jul 2009
Place
Sao Paulo, SP (Brazil)

Optional Information

Notes
Imprint:Anais do 14. Congresso brasileiro de fisica medica; 4. Simposio de instrumentacao e imagens medicas