Published 1983
| Version v1
Journal article
Specific filtration of the primary radiation in computed tomography
Description
Specific filtration with yttrium of the primary radiation on a Somatom 2 whole body scanner improves the relationship absorbed dose/information in terms of contrast resolution while spatial resolution is only marginally affected. (Auth.)
Additional details
Publishing Information
- Journal Title
- Acta Radiol., Diagn.
- Journal Volume
- 24
- Journal Issue
- 1
- Series
- Acta Radiol., Diagn.
- Journal Page Range
- 17-19
- ISSN
- 0567-8056
INIS
- Country of Publication
- Sweden
- Country of Input or Organization
- Sweden
- INIS RN
- 14799143
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- ATTENUATION; CAT SCANNING; ENERGY SPECTRA; EXPERIMENTAL DATA; FILTERS; OXYSULFIDES; PHANTOMS; PHOTONS; RADIATION DOSES; SPATIAL RESOLUTION; X RADIATION; YTTRIUM COMPOUNDS
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DATA; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; INFORMATION; IONIZING RADIATIONS; MASSLESS PARTICLES; MOCKUP; NUMERICAL DATA; OXYGEN COMPOUNDS; RADIATIONS; RESOLUTION; SPECTRA; STRUCTURAL MODELS; SULFUR COMPOUNDS; TOMOGRAPHY; TRANSITION ELEMENT COMPOUNDS