Comparison of two ICP-MS systems for measuring Tc-99 in large volume water samples

  • 1. Risoe National Lab., Dept. of Nuclear Safety Res., Roskilde (Denmark)
  • 2. Risoe National Lab., Dept. of Plant Biology and Biogeochemistry, Roskilde (Denmark)
  • 3. Norwegian Agricultural Univ., Dept. of Chemistry and Biotechnology, Aas (Norway)

Description

Large volume fjord and seawater samples have been radiochemically prepared for ICP-MS analysis in order to test the robustness of the procedure and to carry out a comparison of two ICP-MS systems. A sector field instrument (MicroMass PT2) coupled with an ultrasonic nebuliser and a quadrupole ICP-MS (Perkin Elmer Elan 6000) coupled with an electrothermal vaporisation (ETV) unit were used. The results showed that the radiochemical procedure was robust, removing Ru and Mo to acceptable levels, and that the two instruments gave results that were in agreement. The correlation coefficient between the sets of 11 results was 1.0 ± 0.05. The importance of establishing the matrix effect when using an ETV is discussed. (au)

Part of:
Summaries of studies carried out in the NKS/BOK-2 project. Technical report