Published February 1988 | Version v1
Journal article

Cylindrical pet detector design

  • 1. Physics Research Lab., Massachusetts General Hospital Boston, MA (USA)

Description

A cylindrically shaped high resolution PET detector that uses cross-plane coincidence events is being developed. A 2-D analog coded position sensitive detector is employed. It consists of a hexagonal array of PM Tubes and a rectangular array of crystals, 8 elements per PM Tube. The optics has been designed to maximize the light collection and to provide uniform spatial resolution. The detector will be 60 cm D X 11.5 cm W, the crystals are 3 X 5.7 X 30 mm. The design, associated electronics and results of measurements on a sector of the detector are presented

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
35
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
675-679
ISSN
0018-9499
CODEN
IETNA

Conference

Title
34. nuclear science symposium and 19. nuclear power systems symposium.
Dates
21-23 Oct 1987.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-871006--.