An E/sup 22/B end-loss-ion analyzer for the Tandem Mirror Experiment-Upgrade (TMX-U)

  • 1. Lawrence Livermore National Lab., P.O. Box 5511, L-540, Livermore, CA 94550

Description

The authors have installed a new diagnostic instrument to investigate ions emanating along magnetic-field lines of the TMX-U tandem-mirror experiment. This analyzer contains parallel electric and magnetic fields, which yield ion mass and energy spatial separation. A dual array of 128 copper collector plates detects particles in the ion flux that is first collimated and then focused through the 180-degree bending magnetic field. An electric field applied transverse to the bending particle path then separates the ion masses in the direction perpendicular to the magnetic-pole faces while the magnetic field spreads out the different energies of each mass in a plane parallel to the magnetic-pole tips. The CAMAC-based data recorders are fiber-optically coupled to the system controller for data acquisition, analysis, and display. A commercial CAMAC data recorder was modified for current input. The authors expect to measure higher particle energies than the present gridded end-loss analyzers as well as to more accurately determine the energy spectra

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (USA)
Imprint Title
Proceedings of the 11th symposium on fusion engineering
Journal Page Range
p. 1123-1126.

Conference

Title
11. symposium on engineering problems in fusion research.
Dates
18-22 Nov 1985.
Place
Austin, TX (USA).