Assessment of CaSO4:Dy and LiF:Mg,Ti thermoluminescent dosimeters performance in the dosimetry of clinical electron beams

Description

The assessment of the performance of CaS04:Dy thermoluminescent detectors produced by IPEN in the dosimetry of clinical electron beams aims to propose an alternative to the LiF:Mg,Ti commercial dosimeters (TLD-100) largely applied in radiation therapy. The two types of thermoluminescent dosimeters were characterised with the use of PMMA, RMI-457 type solid water and water phantoms in radiation fields of 4, 6, 9, 12 and 16 MeV electrons of nominal energies in which the dose-response curves were obtained and the surface and depth doses were determined. The thermoluminescent response dependency with the electron nominal energies and the applied phantom were studied. The CaS04:Dy presented the same behaviour than the LiF:Mg,Ti in such a way that its application as an alternative to the TLD-100 pellets in the radiation therapy dosimetry of electron beams is viable and presents the significantly higher sensitivity to the electron radiation as its main advantage. (author)

Availability note (English)

Available from INIS in electronic form

Files

39107608.pdf

Files (4.0 MB)

Name Size Download all
md5:7b8a136303ca21032ae98af5ba058303
4.0 MB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Avaliacao do desempenho dos detectores termoluminesncetes de CaSO

Publishing Information

Imprint Pagination
96 p.
Report number
INIS-BR--4631