Published May 1987 | Version v1
Journal article

Errors in cerebral blood flow determinations by xenon-enhanced computed tomography due to estimation of arterial xenon concentrations

  • 1. Univ. of Pittsburgh, PA

Description

Errors in the determination of xenon concentrations in arterial blood during inhalation of xenon-oxygen mixtures are used to assess errors in the derivation of regional cerebral blood flow by the xenon-enhanced computed tomography (CT) method. The results of this study indicate that approximating the arterial buildup by a single exponential introduces relatively small errors in estimated flow values. The most significant systematic error is introduced by errors in estimation of the xenon arrival time to the brain in relationship to sequential (CT) scanning times

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
14
Journal Issue
3
Series
Med. Phys.
Journal Page Range
377-381
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
19008833
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BLOOD FLOW; BRAIN; DYNAMIC FUNCTION STUDIES; EMISSION COMPUTED TOMOGRAPHY; ERRORS; XENON ISOTOPES
Descriptors DEC
BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; NERVOUS SYSTEM; ORGANS; TOMOGRAPHY