Description
A surface ionization-diffusion-type ionization source that uses a rhenium filament overplated with platinum has been developed and optimized for 0.1-ng neptunium samples. This source is capable of measuring the neptunium content of nuclear-test-debris samples to 0.15% precision at the 95% confidence level. 14 refs., 3 figs., 3 tabs
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE86012746.Files