Published October 2010 | Version v1
Journal article

Characterization of the Li beam probe with a beam profile monitor on JET

  • 1. Instituto de Plasma e FuSao Nuclear, Instituto Superior Tecnico, Associacao EURATOM/IST, 1049-001 Lisboa (Portugal)
  • 2. JET-EFDA, CCFE, Culham Science Centre, Abingdon OX14 3DB (United Kingdom)

Description

The lithium beam probe (LBP) is widely used for measurements of the electron density in the edge plasma of magnetically confined fusion experiments. The quality of LBP data strongly depends on the stability and profile shape of the beam. The main beam parameters are as follows: beam energy, beam intensity, beam profile, beam divergence, and the neutralization efficiency. For improved monitoring of the beam parameters, a beam profile monitor (BPM) from the National Electrostatics Corporation (NEC) has been installed in the Li beam line at JET. In the NEC BPM, a single grounded wire formed into a 45 deg. segment of a helix is rotated by a motor about the axis of the helix. During each full revolution, the wire sweeps twice across the beam to give X and Y profiles. In this paper, we will describe the properties of the JET Li beam as measured with the BPM and demonstrate that it facilitates rapid optimization of the gun performance.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
81
Journal Issue
10
Journal Page Range
p. 10D734-10D734.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2010 American Institute of Physics
Collaborations
JET EFDA Contributors