Published 1998 | Version v1
Journal article

High-performance, imaging, thermal neutron detectors

  • 1. Brookhaven National Laboratory, Instrumentation Division, Building 535B, Upton, NY 11973-5000 (United States)

Description

Existing and planned spallation neutron sources require two-dimensional detectors for many experiments. Unlike the requirements for steady-state neutron sources, it is essential that these detectors possess good time resolution to determine neutron energy. A range of detectors based on gas proportional chambers, with low-noise encoding electronics, has been fabricated at this laboratory, with properties well suited for use at spallation sources. These high-performance detectors possess outstanding qualities in terms of dynamic range and stability of both recorded neutron positions and response (efficiency), in addition to normal attributes such as good position resolution, high detection efficiency and insensitivity to γ-rays. We review here some of the major characteristics of the detectors, describe recent advances, and illustrate their high level of performance with neutron scattering results. While relatively few such detectors are required world wide, specialized efforts are required for their development. The additional opportunities provided by new spallation sources will need continued advances in detector performance. (author)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
419
Journal Issue
Nos.2,3
Journal Page Range
p. 642-647
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
8. International Wire Chamber Conference
Dates
23-27 Feb 1998
Place
Vienna (Austria)