Behavior of a trapezoid-based data acquisition system up to 100 kHz and beyond

  • 1. Goethe University, Max-von-Laue-Strasse 1, 60438 Frankfurt am Main (Germany)
  • 2. Johannes Gutenberg University, Saarstrasse 21, 55122 Mainz (Germany)
  • 3. GSI Helmholtzzentrum für Schwerionenforschung GmbH, Planckstrasse 1, 64291 Darmstadt (Germany)

Description

In this work, we investigated the ability of a high-purity germanium detector connected to a trapezoid-filter-based data acquisition system to reliably record signals in spite of high sample activities. By activating multiple Na2CO3 samples with different Na content, we were able to deduce efficiency, resolution and dead time of the system used as a function of the sample activity. Based on the results, we were able to find a setting which allows measurements of event rates up to 35 kHz per readout channel with an energy resolution of 0.3% at the 2754 keV 24Na line