Accuracy of Robotic Radiosurgical Liver Treatment Throughout the Respiratory Cycle

  • 1. Medical Physics, Juravinski Cancer Center, Hamilton, Ontario (Canada)
  • 2. Department of Oncology, McMaster University, Hamilton, Ontario (Canada)

Description

Purpose: To quantify random uncertainties in robotic radiosurgical treatment of liver lesions with real-time respiratory motion management. Methods and Materials: We conducted a retrospective analysis of 27 liver cancer patients treated with robotic radiosurgery over 118 fractions. The robotic radiosurgical system uses orthogonal x-ray images to determine internal target position and correlates this position with an external surrogate to provide robotic corrections of linear accelerator positioning. Verification and update of this internal–external correlation model was achieved using periodic x-ray images collected throughout treatment. To quantify random uncertainties in targeting, we analyzed logged tracking information and isolated x-ray images collected immediately before beam delivery. For translational correlation errors, we quantified the difference between correlation model–estimated target position and actual position determined by periodic x-ray imaging. To quantify prediction errors, we computed the mean absolute difference between the predicted coordinates and actual modeled position calculated 115 milliseconds later. We estimated overall random uncertainty by quadratically summing correlation, prediction, and end-to-end targeting errors. We also investigated relationships between tracking errors and motion amplitude using linear regression. Results: The 95th percentile absolute correlation errors in each direction were 2.1 mm left–right, 1.8 mm anterior–posterior, 3.3 mm cranio–caudal, and 3.9 mm 3-dimensional radial, whereas 95th percentile absolute radial prediction errors were 0.5 mm. Overall 95th percentile random uncertainty was 4 mm in the radial direction. Prediction errors were strongly correlated with modeled target amplitude (r=0.53-0.66, P<.001), whereas only weak correlations existed for correlation errors. Conclusions: Study results demonstrate that model correlation errors are the primary random source of uncertainty in Cyberknife liver treatment and, unlike prediction errors, are not strongly correlated with target motion amplitude. Aggregate 3-dimensional radial position errors presented here suggest the target will be within 4 mm of the target volume for 95% of the beam delivery.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ijrobp.2015.08.031

Additional details

Identifiers

DOI
10.1016/j.ijrobp.2015.08.031;
PII
S0360-3016(15)03186-7;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
93
Journal Issue
4
Journal Page Range
p. 916-924
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48093614
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; CORRELATIONS; ERRORS; FORECASTING; IMAGES; LINEAR ACCELERATORS; LIVER; POSITIONING; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; BODY; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.