Published April 2010 | Version v1
Journal article

A validation of computational phantoms from photographic images for patient-tailored whole body counting

  • 1. Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)
  • 2. Department of Nuclear Engineering, Hanyang University, Seoul (Korea, Republic of)
  • 3. National Radiation Emergency Medical Center, Korea Institute of Radiological and Medical Science, Seoul (Korea, Republic of)
  • 4. Korea Institute of Nuclear Safety, Daejeon (Korea, Republic of)

Description

This study attempted to validate a new method for patient-tailored efficiency calibration. Digital calibration with Monte Carlo simulations was used to substitute the lack of precision limitation due to the limited number of experimental phantoms in whole body counting calibration for internal dosimetry. The validity of this approach was examined by comparing the simulation results to the measured values from actual measurements using family BOMAB phantoms. The computational voxel phantoms were constructed by a reconstruction technique using AP and lateral photographic images of the BOMAB phantoms, instead of using the given specifications provided with BOMAB phantoms. Although discrepancies to a certain degree between the computational simulation and measured efficiencies do exist, the results support the new approach of being an alternative to family BOMAB phantoms.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2009.10.035

Additional details

Identifiers

DOI
10.1016/j.apradiso.2009.10.035;
PII
S0969-8043(09)00652-6;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
68
Journal Issue
4-5
Journal Page Range
p. 776-779
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
7. international topical meeting on industrial radiation and radio isotope measurement application
Acronym
IRRMA-7
Dates
22-27 Jun 2008
Place
Prague (Czech Republic)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41068565
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; COMPUTERIZED SIMULATION; DOSIMETRY; EFFICIENCY; IMAGES; MONTE CARLO METHOD; PATIENTS; PHANTOMS; VALIDATION; WHOLE-BODY COUNTING
Descriptors DEC
CALCULATION METHODS; COUNTING TECHNIQUES; MOCKUP; SIMULATION; STRUCTURAL MODELS; TESTING