Monte Carlo simulation of scintillator efficiency for GSO single crystal II
Creators
Description
To efficiently utilize light output of GSO scintillator, we have investigated influence of Compton effect, surface roughness, scintillator size and light transmissivity to the energy spectrum using Monte Carlo simulation when irradiated by a γ-ray of 662 keV from 137Cs. The new founding is that Compton effect does not have any influence to the photoelectric peak of the energy spectrum, but the other factors do. When the scintillator is cubic, 10 x 10 x 10 mm3, and all the surfaces are polished, the photoelectric peak follows Gaussian. When the scintillator takes either rough surface or low light transmissivity, the peak tends to deviate from the Gaussian, having a shoulder in the upper side of the peak, which degrades energy resolution. The deviation gets larger when the scintillator is elongated as 5 x 5 x 30 mm3
Additional details
Publishing Information
- Publisher
- IEEE Service Center.
- Imprint Place
- Piscataway, NJ (United States)
- Imprint Title
- 1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
- Imprint Pagination
- 2138 p.
- Journal Page Range
- p. 1219-1222.
Conference
- Title
- Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
- Dates
- 2-9 Nov 1996.
- Place
- Anaheim, CA (United States).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 28068739
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S36: MATERIALS SCIENCE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPTON EFFECT; EFFICIENCY; ENERGY RESOLUTION; GERMANIUM SULFIDES; MONTE CARLO METHOD; POSITRON COMPUTED TOMOGRAPHY; SCINTILLATIONS; SOLID SCINTILLATION DETECTORS
- Descriptors DEC
- BASIC INTERACTIONS; CALCULATION METHODS; CHALCOGENIDES; COMPUTERIZED TOMOGRAPHY; ELASTIC SCATTERING; ELECTROMAGNETIC INTERACTIONS; EMISSION COMPUTED TOMOGRAPHY; GERMANIUM COMPOUNDS; INTERACTIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SCATTERING; SCINTILLATION COUNTERS; SULFIDES; SULFUR COMPOUNDS; TOMOGRAPHY