Equations for Ngas and Nair in terms of NX and NK
Creators
- 1. Department of Medical Physics, University of Wisconsin Medical School, Madison, Wisconsin 53706 (USA)
Description
Task Group 21 of the American Association of Physicists in Medicine defined the cavity-gas calibration factor Ngas for ionization chambers in a protocol for radiotherapy dosimetry published in this journal [Med. Phys. 10, 741 (1983)]. Later, Schulz et al., [Med. Phys. 13, 755 (1986)] published a letter of clarification that proposed a revised equation for Ngas in which the effect of atmospheric humidity was specifically dealt with. The present report points out errors in both of those presentations of the Ngas equation, and derives formulations for cases where the standardization laboratory does or does not correct for ambient humidity during chamber calibration. In the case where humidity is corrected for, the resulting quantity is more properly called ''Nair.'' The Ngas equation is stated in terms of the air kerma calibration factor NK as well as the more familiar exposure calibration factor NX
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 16
- Journal Issue
- 5
- Series
- Med. Phys.
- Journal Page Range
- 803-806
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21020507
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CORRECTIONS; DOSEMETERS; ELECTRON DOSIMETRY; GAMMA DOSIMETRY; HUMIDITY; IONIZATION CHAMBERS; KERMA; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; RADIATION DETECTORS; THERAPY