Published September 1989 | Version v1
Journal article

Equations for Ngas and Nair in terms of NX and NK

Creators

  • 1. Department of Medical Physics, University of Wisconsin Medical School, Madison, Wisconsin 53706 (USA)

Description

Task Group 21 of the American Association of Physicists in Medicine defined the cavity-gas calibration factor Ngas for ionization chambers in a protocol for radiotherapy dosimetry published in this journal [Med. Phys. 10, 741 (1983)]. Later, Schulz et al., [Med. Phys. 13, 755 (1986)] published a letter of clarification that proposed a revised equation for Ngas in which the effect of atmospheric humidity was specifically dealt with. The present report points out errors in both of those presentations of the Ngas equation, and derives formulations for cases where the standardization laboratory does or does not correct for ambient humidity during chamber calibration. In the case where humidity is corrected for, the resulting quantity is more properly called ''Nair.'' The Ngas equation is stated in terms of the air kerma calibration factor NK as well as the more familiar exposure calibration factor NX

Additional details

Publishing Information

Journal Title
Medical Physics
Journal Volume
16
Journal Issue
5
Series
Med. Phys.
Journal Page Range
803-806
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
21020507
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; CORRECTIONS; DOSEMETERS; ELECTRON DOSIMETRY; GAMMA DOSIMETRY; HUMIDITY; IONIZATION CHAMBERS; KERMA; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; RADIATION DETECTORS; THERAPY