A method for measuring the dose distribution of the radiotherapy domain using the computed radiography system
Creators
- 1. Nagoya Univ. (Japan). School of Health Sciences
- 2. Nagoya Univ. (Japan). School of Medicine
Description
Knowing the dose distribution in a tissue is as important as being able to measure exposure or absorbed dose in radiotherapy. Therefore, we have developed a measurement method for the dose distribution (CR dosimetry) in the phantom based on the imaging plate (IP) of the computed radiography (CR). The IP was applied for the dose measurement as a dosimeter instead of the film used for film dosimetry. The data from the irradiated IP were processed by a personal computer with 10 bits and were depicted as absorbed dose distributions in the phantom. The image of the dose distribution was obtained from the CR system using the Digital Imaging and Communications in Medicine (DICOM) form. The CR dosimetry is an application of CR system currently employed in medical examinations to dosimetry in radiotherapy. A dose distribution can be easily shown by the Dose Distribution Depiction System we developed this time. Moreover, the measurement method is simpler and a result is obtained more quickly compared with film dosimetry. (author)
Additional details
Publishing Information
- Journal Title
- Igaku Butsuri
- Journal Volume
- 22
- Journal Issue
- 2
- Journal Page Range
- p. 118-124
- ISSN
- 1345-5354
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 34017566
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPARATIVE EVALUATIONS; COMPUTERS; DEPTH DOSE DISTRIBUTIONS; FILM DOSIMETRY; GAMMA DOSIMETRY; PHANTOMS; RADIATION DOSES; RADIATION MONITORING
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DOSES; DOSIMETRY; EVALUATION; MEDICINE; MOCKUP; MONITORING; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS