Published December 2013
| Version v1
Book
Characterization of liquid scintillators for NAND facility at IUAC
Creators
- 1. Inter University Accelerator Centre, New Delhi (India)
- 2. Physics Department, Punjab University, Chandigarh (India)
Description
The upcoming neutron detector array facility at IUAC will have 100 individual neutron detectors installed in a fixed radius semi sphere configuration. Currently in its first phase of installation, the array consist 50 detectors mounted on a metallic geodesic dome structure at a radial distance of 1.75 m from the target. The detectors are of 5 inch x 5 inch organic liquid scintillator cell BC501-A coupled to 5 inch photo multiplier tube (PMT). The BC501-A liquid organic scintillators have the best pulse shape discrimination properties giving good neutron-gamma discrimination
Additional details
Publishing Information
- Publisher
- Bhabha Atomic Research Centre
- Imprint Place
- Mumbai (India)
- Imprint Title
- DAE-BRNS symposium on nuclear physics. V. 58
- Imprint Pagination
- 1041 p.
- Journal Page Range
- p. 986-987
Conference
- Title
- 58. DAE-BRNS symposium on nuclear physics
- Dates
- 2-6 Dec 2013
- Place
- Mumbai (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 45043267
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- LIQUID SCINTILLATION DETECTORS; NEUTRON DETECTORS; NEUTRON SOURCES; PHOTOMULTIPLIERS; PREAMPLIFIERS; PULSE DISCRIMINATORS
- Descriptors DEC
- AMPLIFIERS; DISCRIMINATORS; ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS; PARTICLE SOURCES; PHOTOTUBES; PULSE CIRCUITS; RADIATION DETECTORS; RADIATION SOURCES; SCINTILLATION COUNTERS
Optional Information
- Notes
- 2 refs., 2 figs.