Published June 1, 2016
| Version v1
Journal article
Calibration Interval Adjustment of a Measuring Instrument in Industries During Long-term Use
Creators
- 1. Associate Professor, National Research Tomsk Polytechnic University, Tomsk (Russian Federation)
- 2. Student, National Research Tomsk Polytechnic University, Tomsk (Russian Federation)
- 3. Senior lecturer, National Research Tomsk Polytechnic University, Tomsk (Russian Federation)
Description
Calibration interval adjustment of measuring instruments is one of the urgent tasks in industries. The article represents the calibration interval calculation of the potentiometer PCB-4P according to the verifications for the 4 year period of the Metrological Department in the aviation plant. The calibration interval is shown to be increased according to the calculation of its reliability and stability of metrological characteristics. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1757-899X/132/1/012029Additional details
Identifiers
Publishing Information
- Journal Title
- IOP Conference Series. Materials Science and Engineering (Online)
- Journal Volume
- 132
- Journal Issue
- 1
- Journal Page Range
- [6 p.]
- ISSN
- 1757-899X
Conference
- Title
- 4. international conference on modern technologies for non-destructive testing
- Dates
- 5-10 Oct 2015
- Place
- Tomsk (Russian Federation)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 50051836
- Subject category
- S42: ENGINEERING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; INDUSTRY; POTENTIOMETERS; RELIABILITY; STABILITY; VERIFICATION
- Descriptors DEC
- ELECTRIC MEASURING INSTRUMENTS; ELECTRICAL EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS