Published January 15, 1988
| Version v1
Journal article
Time resolution of a large area planar spark counter
Creators
- 1. Nara Women's Univ. (Japan). Dept. of Physics
- 2. National Lab. for High Energy Physics, Oho, Ibaraki (Japan)
Description
A planar spark counter of size 1.0 m x 0.1 m has been developed and tested. The constituent elements are a semiconductive glass anode and a window float-cast glass cathode coated by vacuum-deposited copper. The distance between electrodes is 200 μm. A single counter time resolution of 61 ps was achieved with cosmic rays. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 263
- Journal Issue
- 2/3
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 381-386
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 19033447
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANODES; CATHODES; CONSTRUCTION; OPERATION; PERFORMANCE TESTING; SPARK CHAMBERS; TIME RESOLUTION
- Descriptors DEC
- ELECTRODES; GAS TRACK DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; TESTING; TIMING PROPERTIES