Published February 2017
| Version v1
Journal article
Mass test of AdvanSiD model ASD-NUV3S-P SiliconPMs for the Pixel Timing Counter of the MEG II experiment
Creators
- 1. INFN - Sezione di Pavia, via Bassi, 6 – 27100 Pavia (Italy)
Description
The MEG II Timing Counter will measure the positron time of arrival with a resolution of 30 ps relying on two arrays of scintillator pixels read out by 6144 Silicon Photomultipliers (SiPMs) from AdvanSiD. They must be characterized, measuring their breakdown voltage, to assure that the gains of the SiPMs of each pixel are as uniform as possible, to maximize the pixel resolution. To do this an automatic test system that can measure sequentially the parameters of 32 devices has been developed.
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/12/02/C02024Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 12
- Journal Issue
- 02
- Journal Page Range
- p. C02024
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49019989
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BREAKDOWN; ELECTRIC POTENTIAL; EQUIPMENT; MASS; PHOSPHORS; PHOTOMULTIPLIERS; POSITRONS; READOUT SYSTEMS; RESOLUTION; SI SEMICONDUCTOR DETECTORS