Performance of the SLD central drift chamber
Creators
- 1. Massachusetts Inst. of Tech., Cambridge, MA (United States)
- 2. Stanford Linear Accel. Center, CA (United States)
- 3. Victoria Univ., BC (Canada)
- 4. Perugia Univ. (Italy)
- 5. British Columbia Univ., Vancouver, BC (Canada)
- 6. Vanderbilt Univ., Nashville, TN (United States)
Description
The central drift chamber (CDC) of the Stanford Linear Collider Large Detector (SLD) has been recording data since 1992. The low mass of the chamber and the use of a gas characterized by both a low drift velocity and low diffusion constant help to minimize multiple scattering and drift-distance measurement errors. We describe some of the calibrations and corrections applied to the data, and report on the resolutions achieved thus far. We measure an intrinsic drift resolution of 55-110 μm in the region of uniform field. Analysis of the full drift-pulse waveform allows for efficient double-hit resolution of about 1 mm. Transverse momentum resolution is given by the formula (σ(pt)/p2t)2=0.00502+(0.010/pt)2, where pt is in GeV/c. Used in conjunction with the SLD vertex detector, the CDC permits measurements of impact parameters of high-momentum tracks to 10 μm in the r-φ plane and 36 μm in the r-z plane. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 367
- Journal Issue
- 1-3
- Journal Page Range
- p. 111-114.
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 7. international wire chamber conference.
- Dates
- 13-17 Feb 1995.
- Place
- Vienna (Austria).
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 27023076
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COLLIDING BEAMS; CORRECTIONS; DIFFUSION; DRIFT CHAMBERS; ELECTRON DRIFT; ENERGY RESOLUTION; ERRORS; FLAT MAGNETIC SPECTROMETERS; IMPACT PARAMETER; MAGNETIC FIELDS; MINIMIZATION; MULTIPLE SCATTERING; PARTICLE TRACKS; POSITION SENSITIVE DETECTORS; PULSE TECHNIQUES; PULSES; SPATIAL RESOLUTION; STANFORD LINEAR COLLIDER DETECTOR; TRANSVERSE MOMENTUM
- Descriptors DEC
- BEAMS; LINEAR MOMENTUM; MAGNETIC SPECTROMETERS; MEASURING INSTRUMENTS; MULTIWIRE PROPORTIONAL CHAMBERS; OPTIMIZATION; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RESOLUTION; SCATTERING; SPECTROMETERS